I want to write this topic for giving an idea about how to calculate the Assay, related substances and Residual solvents.
Calculations for Assay Method (Tirtimetry)
Titire value(Sample - Blank) x M x Factor x 100 x 100
--------------------------------------------------------
Ws x (100-LOD)
M= Molarity of the volumetric solution
Ws= Weight of the sample
Factor = Factor for Drug substance
LOD is Loss on drying or MC is water content
Calculations for Assay Substances Method (HPLC)
At x Ws x P x (100-WC of std)
Assay % = ----------------------------------
As x Wt x (100-WC of sample)
At = Area due to sample Preparation
As= Area due to working standard solution
Wt= Weight of the sample in mg
Ws= Weight of the working standard
P=Potency or Assay of the standard
WC=% water content of the sample or standard
Note: WC may be replaced by LOD based on the drug substance
Calculations for Related Substances Method (HPLC)
ri
% of Known Impurity = ---x 100 X RF
rs
ri
% of unknown Impurity = ---x 100
rs
Total Impurities = Sum of all known and unknown impurities
ri =Area of each impurity Peak in the chromatogram of the sample solution preparation
rs=Sum of areas of Main drug and all impurity Peaks in the chromatogram of the sample solution preparation
RF= Response factor
Calculations for Residual solvents Method (GC)
At Volume of solvent taken x density 0.1 ml dilution of sample(ml)
= ------x -------------------------------------------x ---------x-------------------------------x 10 ^ 6
As 10 ml 10 ml Weight of the sample
At= Area of the solvent in the sample - Blank
As= Area of the solvents in the standard -blank
Note: Volume of the solvents taken should be mentioned in microliters
As we are multiplying with 10 power 6 we can get the results in PPM
Note: Refer Merck index for solvent densities
This comment has been removed by the author.
ReplyDeleteIn the calculation the total value must be multiplied by 100 for percentage
ReplyDeleteFor Calculation for Assay substances method (By HPLC)
DeleteHello,
DeleteCan you please clarify about the potency..
is the Potency is on 'as-is' basis or 'Anhydrous' basis in the Calculation for Assay...
Hey Chandra,
Deletei think your query makes no sense, as if there is any LOD in standard then there is every chance for Over Assay, so everytime the standard to be dried[if required] till the LOD becomes zero/ until it comes into specified limit.
Regards,
Pharma Engineering
Hello how can i calculate the potency of a peak from a chromatogram. i have an agilent 1100. i have the standards and my sample. Any suggestions?
ReplyDeletespl area/std area*std weight/spl weight*%potency*(100-LOD or water)%*100, in case of tablet put average weight and divide by its label claim * 100
Deleteand in terms of liquid?
Deletecan anybody suggest me why loss on drying or water content of drug substance is not considered while calculating the assay percentage of API in drug product, only standard potency and water or LOD is considered. why is so? please anyone
ReplyDeletecan u give me the formula for calculation of assay of tablet, for example with potency of 98.55 % and water content 10 %, water of sample = 10.25 %, average weight of tablet = 210 mg
ReplyDeleteThis is too good to learn it here, good explanation, thanks for sharing.
ReplyDeleteRegards,
Pharma Engineering
Nice Information and Very Useful for me and also Useful for Who wants more Information About HPLC GC.
ReplyDeletecan you please tell me tge formula for calculating assay of suspension?
ReplyDeleteWt is label claim?
ReplyDeleteVery informative and well written post! Quite interesting and nice topic chosen for the post Nice Post keep it up.Excellent post. I want to thank you for this informative post. I really appreciate sharing this great post. Keep up your work.
ReplyDeleteassay sample
How can we calculate assay in liquids dosage forms which contain wt/ml??
ReplyDelete